Introduction
Message Attributes
Message Values
Configuration
Starting the MESA Servers
Loading Test Data
Evaluation Scripts
Submission of Results
Individual Tests for Diagnostic Workstations
Individual Tests for Measurement Workstations
Individual Tests for Cardiology
We invent the term Diagnostic Reporting Station for use in this document. This is not an IHE term, but we use it to describe workstations that are used to create reports in the context of the user reviewing other evidence first (such as images). This term will be used to designate tests that are mandatory for your system. This does not imply that your system must display images.
We use the term Measurement Station to describe a device that creates original data (such as numeric measurements in an Ultrasound report). Again, this is not an IHE term but is needed to document test requirements.
If your Report Creator is not described by either category, you should contact the Technical Project Manager and describe your application. This will allow us to construct reasonable tests for your system.
For more details on this template, please reference Supplement 53: DICOM Content Mapping Resource.
This document lists a number of tests for Report Creator Systems. You may not be responsible for all of these tests.
Please refer to the Connectathon web tool to list the required tests for your system. The web address of this tool depends on the year and project manager. Please contact the appropriate project manager to obtain this information.
DICOM SR Objects contain attributes that are not used or required by the IHE Technical Framework. This section contains a table defining those attributes that will be evaluated in the Report Creator tests. Other values may be required by the Technical Framework but not evaluated with this software.
In the table below, the columns marked Value and Present indicate the scope of the evaluation for these messages. Fields that are marked in the Value column indicate the tests will expect a specific value in the field (for example, we specify the patient name). Fields marked in the Present column are tested to see if the field contains a value, but the tests cannot evaluate for a specific value.
Tag |
Attribute Name |
Value |
Present |
0008 0016 |
SOP Class UID |
x |
|
0008 0050 |
Accession Number |
x |
|
0008 0060 |
Modality |
x |
|
0010 0010 |
Patient Name |
x |
|
0010 0020 |
Patient ID |
x |
|
0020 000D |
Study Instance UID |
x |
|
0040 A040 |
Value Type |
x |
|
0040 A043 |
Concept Name Code Sequence (Report Title) |
|
|
> 0008 0100 |
Code Value |
x |
|
> 0008 0102 |
Coding Scheme Designator |
x |
|
> 0008 0104 |
Code Meaning |
x |
|
0040 A491 |
Completion Flag |
x |
|
0040 A493 |
Verification Flag |
x |
|
0040 A504 |
Content Template Sequence |
|
|
> 0008 0101 |
Mapping Resource |
x |
|
> 0040 DB00 |
Template Identifier |
x |
|
0040 A730 |
Content Sequence |
|
|
> 0040 A010 |
Rerlationship Type |
x |
|
> 0040 A040 |
Relationship Value |
x |
|
Other attributes |
As needed for the Relationship Value |
x |
|
The table below gives parameters for MESA servers that will receive messages from your system.
Application |
AE Title / Receiving Application |
Port Number |
MESA Image Manager |
MESA_IMG_MGR |
2350 |
MESA Report Manager |
REPORT_MANAGER |
2700 |
MESA Audit Record Repository |
<none> |
4000 |
MESA Report Manager for HL7 message |
MESA_RPTMGR |
2750 |
This section details configuration information if the Report Creator under test is coupled to an Image Display. It describes configuration for the Image Display.
The MESA Image Manager maintains a database of DICOM applications used for C-Move operations.
Add an entry for the storage SCP associated with your workstation. Edit the text file $MESA_TARGET/db/loaddicomapp.pgsql (Unix) or $MESA_TARGET/db/loaddicomapp.sql (Windows 2000 or XP) Use the existing entries as a template and add entries for your workstations as appropriate. The column names found in the SQL insert statements are described in the following table.
Column Name |
Description |
aet |
DICOM Application Entity Title. Must be unique. |
host |
Host name (or IP address) of the application. |
port |
TCP/IP port number for receiving associations. |
org |
The organization that operates the device. Useful if multiple organizations use the Image Manager. |
com |
A comment field. |
You can test your work as follows:
perl load_apps.pl imgmgr
These instructions assume you are using a terminal emulator on Unix systems or an MS DOS command window under Windows 2000/XP. Each test uses a command line interface; there is no graphical user interface. Before you start the test procedure, you need to start the MESA Order Placer and MESA Order Filler servers. Make sure the appropriate database is running (PostgreSQL, SQL Server). To start the MESA servers:
scripts/start_mesa_servers.csh (Unix)
scripts\start_mesa_servers.bat (Windows)Log levels are set for the MESA Image Manager in the file: $MESA_TARGET/runtime/rpt_manager/ds_dcm.cfg. / Log levels are:
When you are finished running one or more tests, you can stop the servers:
scripts/stop_mesa_servers.csh (Unix)
scripts\stop_mesa_servers.bat (Windows)Log files are stored in $MESA_TARGET/logs.
Some Report Creator tests will make use of images retrieved from the MESA Image Manager. If the Report Creator under test is coupled with an Image Display, you should load test images into the MESA Image Manager.
perl 60x/load_img_mgr.pl
The test and evaluation scripts are modified to take a <log> parameter. This number indicates how much logging to include in the output. Defined values are listed in the table below.
Level |
Meaning |
1 |
Errors only |
2 |
Warnings + level 1 |
3 |
Verbose mode + level 2 |
4 |
References in Technical Framework + level 3 |
Not all test and evaluation scripts have been updated as of this release. When you run the scripts with no arguments, they will tell you what the real arguments are. Each test description that follows in this document will be updated correctly when the test script is updated.
Test descriptions below inform the reader to “submit results to the Project Manager”. This is does not mean “email”. The current submission process should be documented by the Project Manager, but will not include emailing files directly to the Project Manager.
Test 600: Example SR ObjectIn test 600, the Report Creator creates a sample SR object and submits it to the Project Manager. The purpose is to distribute the SR document to other participants. ReferencesInstructions
EvaluationSupplemental Information |
Test 601 is required of Diagnostic Reporting Stations. This test will require you to create a structured report according to the Basic Diagnostic Imaging Report template (TID 2000). All items in the template that are marked mandatory will be tested. In addition, values in the following tables will also be required. Items not specified in the tables may be assigned any value. Table 2.1-1 lists the header attributes that will be tested. Report Creators may be standalone or coupled with an Image Display. Standalone Report Creators are expected to fill in most attributes found in Table 2.1-1, but not all. The attributes marked with x in the Standalone or w/Image Display columns indicate those values for which the tests expect an exact match. For example, a Report Creator coupled with an Image Display is expected to reuse the Study Instance UID from the images and not create a new study for this report.
Table 2.1-1 Selected Attributes tested in Test 601
Attribute Name |
Tag |
Standalone |
w/Image Display |
Patient Name |
0010 0010 |
x |
x |
Patient ID |
0010 0020 |
x |
x |
Patient Sex |
0010 0040 |
x |
x |
Accession Number |
0008 0050 |
x |
x |
Requested Procedure ID |
|
|
x |
Study Instance UID |
0020 000D |
|
x |
Completion Flag |
0040 A491 |
x |
x |
Verification Flag |
0040 A493 |
x |
x |
The 601 suite of tests require you to create a report according to TID2000, but we specify observation context and three sections of the report. Figure 2-1 below shows the overall structure and relationship of content items required for the 601 tests. You are expected to create a report with the section headings as defined in the tables found in the following sections. We do not require the text in the sections to match the text in our samples. The text is only provided as an example.

Figure 2-1 Content Item Requirements for Test 601
The 601 tests are broken down further by imaging modality: CR, CT, or MR. If your Report Creator is not coupled with an Image Display, run test 601. If you are coupled with an Image Display, run test 601CR, 601CT, or 601MR as appropriate. You may run all four tests, but are only required to submit results for one. Values for content items (listed in Value Set Constraint) that are italicized are suggested for each modality, and not required. That is, the software tests report headings but not the exact text in a section of the report.
perl scripts/reset_rpt_mgr.pl
perl 601/eval_601.pl
Attribute |
Value |
Patient Name |
CRTHREE^PAUL |
Patient ID |
CR3 |
Patient Sex |
M |
Accession Number |
2001C30 |
Completion Flag |
COMPLETE |
Verification Flag |
UNVERIFIED |
|
NL |
Rel with Parent |
VT |
Concept Name |
Value Set Constraint |
1 |
|
|
CONTAINER |
BCID (7000) Diagnostic Imaging Report Document Titles |
(LN, 11528-7, “Radiology Report”) |
2 |
> |
HAS CONCEPT MOD |
CODE |
(121049,DCM,”Language of Content Item and Descendents”) |
(ISO639_2, eng, English) |
3 |
> |
HAS OBS CONTEXT |
CODE |
(DCM, 121005, “Observer Type”) |
(DCM, 121006, Person) |
4 |
> |
HAS OBS |
PNAME |
(DCM, 121008, “Person Observer Name”) |
MOORE^STEVE |
5 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121060, History) |
6 |
>> |
CONTAINS |
TEXT |
(DCM, 121060, History) |
“Cough” |
7 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121070, Findings) |
8 |
>> |
CONTAINS |
TEXT |
(DCM, 121071, Finding) |
“PA evaluation of the chest demonstrates the lungs to be expanded and clear.” |
9 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121076, Conclusions) |
10 |
>> |
CONTAINS |
TEXT |
(DCM, 121077, Conclusion) |
“Normal PA chest x-ray.” |
perl scripts/reset_rpt_mgr.pl
perl 601/eval_601_cr.pl
Attribute |
Value |
Patient Name |
CRTHREE^PAUL |
Patient ID |
CR3 |
Patient Sex |
M |
Accession Number |
2001C30 |
Study Instance UID |
1.113654.1.2001.30 |
Completion Flag |
COMPLETE |
Verification Flag |
UNVERIFIED |
|
NL |
Rel with Parent |
VT |
Concept Name |
Value Set Constraint |
1 |
|
|
CONTAINER |
BCID (7000) Diagnostic Imaging Report Document Titles |
(LN, 11528-7, “Radiology Report”) |
2 |
> |
HAS CONCEPT MOD |
CODE |
(121049,DCM,”Language of Content Item and Descendents”) |
(ISO639_2, eng, English) |
3 |
> |
HAS OBS CONTEXT |
CODE |
(DCM, 121005, “Observer Type”) |
(DCM, 121006, Person) |
4 |
> |
HAS OBS |
PNAME |
(DCM, 121008, “Person Observer Name”) |
MOORE^STEVE |
5 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121060, History) |
6 |
>> |
CONTAINS |
TEXT |
(DCM, 121060, History) |
“Cough” |
7 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121070, Findings) |
8 |
>> |
CONTAINS |
TEXT |
(DCM, 121071, Finding) |
“PA evaluation of the chest demonstrates the lungs to be expanded and clear.” |
9 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121076, Conclusions) |
10 |
>> |
CONTAINS |
TEXT |
(DCM, 121077, Conclusion) |
“Normal PA chest x-ray.” |
perl scripts/reset_rpt_mgr.pl
perl 601/eval_601_ct.pl
Attribute |
Value |
Patient Name |
CTFIVE^JIM |
Patient ID |
CT5 |
Patient Sex |
M |
Accession Number |
2001B20 |
Study Instance UID |
1.113654.1.2001.20 |
Completion Flag |
COMPLETE |
Verification Flag |
UNVERIFIED |
|
NL |
Rel with Parent |
VT |
Concept Name |
Value Set Constraint |
1 |
|
|
CONTAINER |
BCID (7000) Diagnostic Imaging Report Document Titles |
(LN, 18747-6, “CT Report”) |
2 |
> |
HAS CONCEPT MOD |
CODE |
(121049,DCM,”Language of Content Item and Descendents”) |
(ISO639_2, eng, English) |
3 |
> |
HAS OBS CONTEXT |
CODE |
(DCM, 121005, “Observer Type”) |
(DCM, 121006, Person) |
4 |
> |
HAS OBS |
PNAME |
(DCM, 121008, “Person Observer Name”) |
MOORE^STEVE |
5 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121060, History) |
6 |
>> |
CONTAINS |
TEXT |
(DCM, 121060, History) |
“46-year-old male presented to the emergency room with parasternal and epigastric pain. ” |
7 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121070, Findings) |
8 |
>> |
CONTAINS |
TEXT |
(DCM, 121071, Finding) |
“There is an aneurysm of the entire thoracic aorta measuring up to 6.2 cm in diameter. ” |
9 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121076, Conclusions) |
10 |
>> |
CONTAINS |
TEXT |
(DCM, 121077, Conclusion) |
“Aneurysm of thoracic aorta with dissection affecting ascending and descending aorta.” |
perl scripts/reset_rpt_mgr.pl
perl 601/eval_601_mr.pl
Attribute |
Value |
Patient Name |
MRTHREE^STEVE |
Patient ID |
MR3 |
Patient Sex |
M |
Accession Number |
2001A10 |
Study Instance UID |
1.113654.1.2001.10 |
Completion Flag |
COMPLETE |
Verification Flag |
UNVERIFIED |
|
NL |
Rel with Parent |
VT |
Concept Name |
Value Set Constraint |
1 |
|
|
CONTAINER |
BCID (7000) Diagnostic Imaging Report Document Titles |
(LN, 18755-9, “MRI Report”) |
2 |
> |
HAS CONCEPT MOD |
CODE |
(121049,DCM,”Language of Content Item and Descendents”) |
(ISO639_2, eng, English) |
3 |
> |
HAS OBS CONTEXT |
CODE |
(DCM, 121005, “Observer Type”) |
(DCM, 121006, Person) |
4 |
> |
HAS OBS |
PNAME |
(DCM, 121008, “Person Observer Name”) |
MOORE^STEVE |
5 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121060, History) |
6 |
>> |
CONTAINS |
TEXT |
(DCM, 121060, History) |
“Rule out internal derangement.” |
7 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121070, Findings) |
8 |
>> |
CONTAINS |
TEXT |
(DCM, 121071, Finding) |
“Within the posterior horn of the medial meniscus there is minimal increase in signal intensity.” |
9 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121076, Conclusions) |
10 |
>> |
CONTAINS |
TEXT |
(DCM, 121077, Conclusion) |
“Joint diffusion with superior joint plicae.” |
Report Creator Test 602: Simple Imaging Report with Image ReferenceTest 602 is required of Diagnostic Reporting Stations that are coupled with an Image Display. This test uses the same format as Report Creator Test 601. This test requires the Report Creator to add a reference to an image in the SR document.
The 602 tests are broken down by imaging modality: CR, CT, or MR. Run test 602CR, 602CT, or 602MR as appropriate. You may run all three tests, but are only required to submit results for one. Values for content items (listed in Value Set Constraint) that are italicized are suggested for each modality, and not required. That is, the software tests report headings but not the exact text in a section of the report. Report Creator Test 602: CR ImagesThe CR study has 3 images in one series. The original image is a normal chest image. The other two images are derived from the original image and only show the left half and right half. The SR report should use the original (full) image. ReferencesInstructions
Evaluation
Supplemental InformationThe tables below list the demographics for the patient for this test and the coded values for the sections in the DICOM SR to be created.
Report Creator Test 602: CT ImagesReferencesInstructions
Evaluation
Supplemental InformationThe tables below list the demographics for the patient for this test and the coded values for the sections in the DICOM SR to be created.
Report Creator Test 602: MR ImagesReferencesInstructions
Evaluation
Supplemental InformationThe tables below list the demographics for the patient for this test and the coded values for the sections in the DICOM SR to be created.
|
Test 603 is required of Diagnostic Reporting Stations that are coupled with an Image Display. This test uses the same format as Report Creator Test 601. This test requires the Report Creator to add references to two images in the SR document. This is quite similar to the 602 test.
Table 2.1-1 lists the header attributes that will be tested. Report Creators may be standalone or coupled with an Image Display.
The 603 suite of tests require you to create a report according to TID2000, but we specify observation context and three sections of the report.
Figure 2-3 below shows the overall structure and relationship of content items required for the 603 tests.
You are expected to create a report with the section headings as defined in the tables found in the following sections.
We do not require the text in the sections to match the text in our samples.
The text is only provided as an example.
The tests require you to reference two specific images in the study. The tables below will have an entry like the example show below.
The rightmost column indicates the image that should be used as a reference.
We use the triplet (accession number / series number / image number).
In the example below, the series number happens to be blank because that is how the original image is formatted.
9 |
>> |
CONTAINS |
IMAGE |
(DCM, 121080, “Best illustration of findings”) |
2001C30 / / 40 |

Figure 2-3 Content Item Requirements for Test 603
The 603 tests are broken down by imaging modality: CR, CT, or MR. Run test 603CR, 603CT, or 603MR as appropriate. You may run all three tests, but are only required to submit results for one. Values for content items (listed in Value Set Constraint) that are italicized are suggested for each modality, and not required. That is, the software tests report headings but not the exact text in a section of the report.
The CR study has 3 images in one series. The original image is a normal chest image. The other two images are derived from the original image and only show the left half and right half. The SR report should use the two derived images (left view, right view).
perl scripts/reset_rpt_mgr.pl
perl 603/eval_603_cr.pl
Attribute |
Value |
Patient Name |
CRTHREE^PAUL |
Patient ID |
CR3 |
Patient Sex |
M |
Accession Number |
2001C30 |
Study Instance UID |
1.113654.1.2001.30 |
Completion Flag |
COMPLETE |
Verification Flag |
UNVERIFIED |
|
NL |
Rel with Parent |
VT |
Concept Name |
Value Set Constraint |
1 |
|
|
CONTAINER |
BCID (7000) Diagnostic Imaging Report Document Titles |
(LN, 11528-7, “Radiology Report”) |
2 |
> |
HAS CONCEPT MOD |
CODE |
(121049,DCM,”Language of Content Item and Descendents”) |
(ISO639_2, eng, English) |
3 |
> |
HAS OBS CONTEXT |
CODE |
(DCM, 121005, “Observer Type”) |
(DCM, 121006, Person) |
4 |
> |
HAS OBS |
PNAME |
(DCM, 121008, “Person Observer Name”) |
MOORE^STEVE |
5 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121060, History) |
6 |
>> |
CONTAINS |
TEXT |
(DCM, 121060, History) |
“Cough” |
7 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121070, Findings) |
8 |
>> |
CONTAINS |
TEXT |
(DCM, 121071, Finding) |
“PA evaluation of the chest demonstrates the lungs to be expanded and clear.” |
9 |
>> |
CONTAINS |
IMAGE |
(DCM, 121080, “Best illustration of findings”) |
2001C30 / / 40 |
10 |
>> |
CONTAINS |
IMAGE |
(DCM, 121080, “Best illustration of findings”) |
2001C30 / / 60 |
11 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121076, Conclusions) |
12 |
>> |
CONTAINS |
TEXT |
(DCM, 121077, Conclusion) |
“Normal PA chest x-ray.” |
perl scripts/reset_rpt_mgr.pl
perl 603/eval_603_ct.pl
Attribute |
Value |
Patient Name |
CTFIVE^JIM |
Patient ID |
CT5 |
Patient Sex |
M |
Accession Number |
2001B20 |
Study Instance UID |
1.113654.1.2001.20 |
Completion Flag |
COMPLETE |
Verification Flag |
UNVERIFIED |
|
NL |
Rel with Parent |
VT |
Concept Name |
Value Set Constraint |
1 |
|
|
CONTAINER |
BCID (7000) Diagnostic Imaging Report Document Titles |
(LN, 18747-6, “CT Report”) |
2 |
> |
HAS CONCEPT MOD |
CODE |
(121049,DCM,”Language of Content Item and Descendents”) |
(ISO639_2, eng, English) |
3 |
> |
HAS OBS CONTEXT |
CODE |
(DCM, 121005, “Observer Type”) |
(DCM, 121006, Person) |
4 |
> |
HAS OBS |
PNAME |
(DCM, 121008, “Person Observer Name”) |
MOORE^STEVE |
5 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121060, History) |
6 |
>> |
CONTAINS |
TEXT |
(DCM, 121060, History) |
“46-year-old male presented to the emergency room with parasternal and epigastric pain. ” |
7 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121070, Findings) |
8 |
>> |
CONTAINS |
TEXT |
(DCM, 121071, Finding) |
“There is an aneurysm of the entire thoracic aorta measuring up to 6.2 cm in diameter. ” |
9 |
>> |
CONTAINS |
IMAGE |
(DCM, 121080, “Best illustration of findings”) |
2001B20 / 1 / 66 |
10 |
>> |
CONTAINS |
IMAGE |
(DCM, 121080, “Best illustration of findings”) |
2001B20 / 1 / 67 |
11 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121076, Conclusions) |
12 |
>> |
CONTAINS |
TEXT |
(DCM, 121077, Conclusion) |
“Aneurysm of thoracic aorta with dissection affecting ascending and descending aorta.” |
perl scripts/reset_rpt_mgr.pl
perl 602/eval_602_mr.pl
Attribute |
Value |
Patient Name |
MRTHREE^STEVE |
Patient ID |
MR3 |
Patient Sex |
M |
Accession Number |
2001A10 |
Study Instance UID |
1.113654.1.2001.10 |
Completion Flag |
COMPLETE |
Verification Flag |
UNVERIFIED |
|
NL |
Rel with Parent |
VT |
Concept Name |
Value Set Constraint |
1 |
|
|
CONTAINER |
BCID (7000) Diagnostic Imaging Report Document Titles |
(LN, 18755-9, “MRI Report”) |
2 |
> |
HAS CONCEPT MOD |
CODE |
(121049,DCM,”Language of Content Item and Descendents”) |
(ISO639_2, eng, English) |
3 |
> |
HAS OBS CONTEXT |
CODE |
(DCM, 121005, “Observer Type”) |
(DCM, 121006, Person) |
4 |
> |
HAS OBS |
PNAME |
(DCM, 121008, “Person Observer Name”) |
MOORE^STEVE |
5 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121060, History) |
6 |
>> |
CONTAINS |
TEXT |
(DCM, 121060, History) |
“Rule out internal derangement.” |
7 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121070, Findings) |
8 |
>> |
CONTAINS |
TEXT |
(DCM, 121071, Finding) |
“Within the posterior horn of the medial meniscus there is minimal increase in signal intensity.” |
9 |
>> |
CONTAINS |
IMAGE |
(DCM, 121080, “Best illustration of findings”) |
2001A10 / 103 / 45 |
10 |
>> |
CONTAINS |
IMAGE |
(DCM, 121080, “Best illustration of findings”) |
2001A10 / 103 / 46 |
11 |
> |
CONTAINS |
CONTAINER |
BCID (7001) Diagnostic Imaging Report Headings |
(DCM, 121076, Conclusions) |
12 |
>> |
CONTAINS |
TEXT |
(DCM, 121077, Conclusion) |
“Joint diffusion with superior joint plicae.” |
Report Creator Test 604: Partial and UnverifiedThis test is identical to the 601 tests, except for the values of document flags in the SR objects. The Completion Flag (0040 A491) should be set to PARTIAL, and the Verification Flag (0040 A493) should be set to UNVERIFIED. |
This test is identical to the 601 tests, except for the values of document flags in the SR objects. The Completion Flag (0040 A491) should be set to PARTIAL, and the Verification Flag (0040 A493) should be set to VERIFIED.
Report Creator Test 606: Complete and VerifiedThis test is identical to the 601 tests, except for the values of document flags in the SR objects. The Completion Flag (0040 A491) should be set to COMPLETE, and the Verification Flag (0040 A493) should be set to VERIFIED. |
This test is identical to Test 601 with the addtion of the Predecessor Documents Sequence. The goal of this test is to create one report and then to create a second, amended report. You will send both reports to the MESA Report Manager. Follow the instructions for the 601 test, but send both the original and the amended report to the MESA Report Manager. Remember to clear the Report Manager if you need to start the test over.
After the second report has been transmitted, run the evaluation script as appropriate for your system. In addition to the attributes tested in the 601 tests, we evaluate the contents of the Predecessor Documents Sequence (0040 A360) found in your second report.
These tests are for functionality that might not be included with your application. This is an optional test.
Report Creator Test 651: SINR for Measurement WorkstationsTest 651 is required of Measurement Workstations. This test will require you to create a structured report according to the Basic Diagnostic Imaging Report template (TID 2000). All items in the template that are marked mandatory will be tested. ReferencesInstructions
Evaluation
Supplemental Information
|
Report Creator Test 1601: Interpretation WorklistTest 1601 covers the Interpretation task in the Reporting Workflow integration profile (see IHE TF, Vol I: section 13.3). This test only covers the interpretation step and does not include creation of any other items (such as transcriptions or reports). ReferencesInstructionsTo complete this test, your Report Creator needs to be able to:
To run this test: perl scripts/rptcrt_rwf.pl 1601 <log> EvaluationTo evaluate this test: perl 1601/eval_1601.pl <log> <GPPS AE title>Where log is the amount of logging : |
This section describes tests that are specific to the IHE Basic Security integration profile. If you have the MESA servers running for the “standard” tests, you should stop those servers now. You will need to start the MESA secure servers with a different script.
Report Creator Test 1511: Simple Imaging ReportReport Creator Test 1511 uses the same structure as test 601. The Report Creator is expected to communicate with other systems using TLS negotiation and to send appropriate audit messages to the MESA syslog server.
ReferencesInstructions
Evaluation
Supplemental Information
|
Report Creator Test 20501: Verify HL7 Encapsulated Report Message Content (CARD-7)Test 20501 tests the creation and content of an HL7 message with an encapsulated PDF report. ReferencesInstructionsTo run this test, follow these steps using a DOS window or terminal emulator:
Evaluation
Supplemental InformationFor this test, the MESA tools evaluate the following elements:
For OBX w/ Study Instance UID, verify:
For OBX w/ PDF report, verify:
Any missing or incorrect elements will result in a “failure”. You cannot ignore this failure. |
Test 20502 is the simple visual verification of the PDF content. This test tests the case where the PDF becomes disassociated with the PDF message content.
In this test there is no MESA software or scripts involved. Any patient name and ID can be used.
Using a “good” report within your Report Creator, the vendor should create a word or pdf file using the document naming convention of: CompanyName_Product_ 20502_RC_2005.doc.
Using any vendor report viewing tool cut and paste in information from your own product which demonstrates that the followng data elements are visible in a report:
The following items are recommended, but not required (please note which are included in an email):
Test Case 20510: Describe Report Creator MethodsDo not divulge any proprietary information. Create a Word document with the following naming convention: CompanyName_Product_20510_RC_2005.doc and submit it to the Project Manager.
|